CAMERA & DETECTORS

Light Radiometer

Lab Radiometer – ILT5000

The ILT5000 is “The ILT1700 for the 21st Century” improving on the industry standard with rapid measurements (up to 100 Hz), a broader dynamic range (100fA to 1mA), extensive supporting software apps, wireless coomunication, internal data storage, and 4-20 mA output.

The broad linear dynamic range of the ILT5000 and the SMA input connector allow the meter to also serve as a broad range, highly sensitive and accurate range pico ammeter. The ILT5000 supports numerous light measurement applications including, but not limited to; Radiometry, Photometry, Reasearch, UVGI- Sterilization, Solar, Photoresist, Optical Radiation Hazard, Phototherapy, Photo-degradation, Plant Growth and more.

Features
  • 10 Decade Dynamic Range of Optical Analysis
  • Wireless Transmitter, Built-in
  • Made in USA
  • Autorange, Autodark, Auto-sample with User Control Options
  • 4-20 mA Output
  • “Set It & Forget It” Remote Data Logging
  • Built-in Re-Chargeable Battery Pack
  • Multi-System Simultaneous Continuous Monitoring
  • DataLight II Complementary Light Measurement and Data Collection Software Apps
  • Labview Compatible
  • NIST-Traceable/ISO17025 Accredited Calibration and Certificate
  • Backwards Compatible with ILT1700 SED Detector/Filter/Optics
  • Sample Rate Up to 100 Hz, Programmable
  • Custom & OEM Inquiries Welcome
Key Applications
  • Radiometry
  • Photometry
  • Reasearch, UVGI-Sterilization
  • Solar, Photoresist
  • Optical Radiation Hazard
  • Phototherapy
  • Photo-degradation
  • Plant Growth and more
Key Specifications
MeterMeasurement Range100 fA – 1 mA
Reverse Bias2 Voltages/selectable
Automatic RangingRapid switching through all current levels
WiFiEnabled
Operating Temperature-40 to 85°C (Calibrated Irradiance 0-40°C)
USBUSB2, Including Power, for Single and Multiple Systems
USB Current Draw500 mA max, 250 mA typical
Size1-3/5” H x 5” W x 7” L
Input connectorD-Sub and/or SMA
CE certifiedNo rf/noise
SoftwareCLICommand line interface sent over a serial port
TriggerTTL, for synchronization of measurement
Data StorageInternal (includes units, date and time stamps)
Time LineScheduled data acquisition
Integration RangeUnlimited dosage/exposure
Transmission/Reflectance100%
Optical Density0.00 to 4.00
ZeroDark or ambient subtraction
Direct ReadingCurrent or calibrated optical units
AppsMeters, CLI, Bar, Trend and Reports apps all included
Minimum Computer System RequirementsTabletMinimum 8” tablet
Operating SystemWindows 7 or 8
Memory/RAMMinium 1 GB RAM (on board)
Storage32GB (flash or SSD)
Processorat least1.83GHz (with minimum 2MB Cache)
Ports1 Micro-USB 2.0 Port
Wi-Fi® Wireless networking802.11b/g/n